48 RL-TA Series Operation Manual

Hints

If more than one weight is used in combination, enter the total error of the weights to be used.

The range of the weight error should be within

±100.00 mg. If any value outside the range is entered, it

will result in the display of [

r - Err]. When [r - Err] is displayed, press any key to return the display

to Step 2.

14.7 Advice CAL and Full-automatic Span Adjustment

When Advice CAL determines that a certain amount of time has elapsed or that the temperature has

changed since the previous span adjustment, the “CAL” sign flashes, indicating that the balance needs to

be calibrated. Perform a span adjustment immediately when the “CAL” sign begins flashing.

To use this function, set Function 1 to [

E. AdC 1].

“Full-automatic span adjustment” is available only with the HTR series.

This function automatically performs a span adjustment with built-in weights when “Advice CAL”

determines that the balance needs to be calibrated, causing the “CAL” sign to flash.

The span adjustment using built-in weights will be performed about 10 minutes after the [CAL] sign

begins flashing, once the balance has stabilized with no load.

“Full-automatic span adjustment” can be performed only when [

8. CA. 1] (span adjustment with built-in

weights) is enabled.

14.8 Auto Repeatability Measurement (ARM)

This is a function to calculate a standard deviation by loading and unloading the built-in weight and

repeating span adjustment 10 times. Stable measurement can be performed by adjusting the response

through the single-touch response setting function.


Press and hold the Cal key for a few seconds. When

[

ArN] is displayed, release the key.

After a while, the display switches to [

Cnt. 1] and the

measurement starts. As the repeatability measurement

proceeds, the number of measurements displayed in

[

Cnt. *] increases accordingly until it reaches [10].

When the measurement is completed, the standard deviation, for example, [

0.000073g], is displayed.

Pressing any key returns the display to measurement mode.


To cancel the measurement, press the Print key.

Refer to page 12 for setting the single-touch response.