Operation 103

Split Screen Mode

Available on HI 6310 Only

Before wiring and mounting instruments to be displayed on split screen

Be sure to update the firmware in the Display and each instrument to the latest version to
ensure all menus are up to date. The display must only be connected to a single instrument
to upload new firmware.

Wiring the Units to the Display Panel

Wire the instruments together and connect them to the display panel connector wire.

A sample wiring diagram for wiring four HI 6310 instruments to one HI 6110 Panel

Making Longer Display Interface Cables

If you need to install the front panel further away from the HI 6300 series instruments than
is possible with the supplied cable then, you will need to make a display cable between the
panel and the first instrument. in the series of up to four connected units.

Wiring Specifications

Wire size: 20 AVG maximum / 26 AWG minimum

Maximum cable length: 100 ft. (30.48 meters)

Use three twisted pairs with a drain wires

Pair wires +12 and GND, D1 and D2, D3 and D4