Instrument Configuration

57

Hardy Port Parameter

The Hardy Port parameter provides the service port which is combined with the IP address
to provide a unique application socket.

The Hardy Port value can be any 16-bit value between 0 and 65,535.

On the Web page, select the communication menu and the Ethernet TCP/IP submenu, and
left click inside the Hardy Port text field and enter the Hardy Port value.

Save the value by clicking on the SAVE PARAMETERS button

On the HI 6300 series instrument display, select the communication menu and use the UP
or DOWN button to select the Ethernet UDP submenu, then select the Hardy Port menu
item. The current Hardy Port value will be displayed, press the ENTER or the RIGHT
button if the value needs to be modified.

Modbus-RTU Commands and Parameters

Slave Address Parameter

The Slave Address parameter is a unique network address between 1 and 247

assigned to the HI 6300 series instrument.

On the Web page, select the communication menu and then select the Modbus RTU

submenu, and left click inside the Slave Address text field and enter the Slave
Address assigned to the HI 6300 series instrument.

Parameter Name

Parameter ID

Tare command

0x07FF

Zero command

0x0800

Cal Low command

0x0801

Cal High command

0x0802

C2 Cal command

0x0803

Save command

0x319F