Cal tolerance parameter – Hardy HI 6300 Series User Manual

Page 59

Advertising
background image

Instrument Configuration

51

Cal Tolerance Parameter

The value you enter for Calibration Tolerance sets the amount of deviation the instrument
will allow during the calibration process. This value must be greater than or equal to the
base motion value and/or the Graduation Sizes.

The base motion value can be calculated by using the following formula:

Base Motion value = (Total Load Cell Capacity x 0.0003)

We recommend three graduation sizes.

RANGE: 000001 - 999999 (default 10)

On the configuration Web page, enter the value in the text box provided.

On the HI 6300 series instrument display, select the calibration menu and use the UP or

DOWN button to select the Hard Cal submenu, then select the Cal Tolerance (Calibration
Tolerance) menu item. The current value will be displayed, press the ENTER or the RIGHT
button if this value needs to be modified.

Do Cal Lo Command

Before executing this Do Cal Lo command verify that the Cal Lo Weight value is

correct.

After setting/verifying the Cal Lo Weight value executing this command will run the

Hard Calibration process using the Cal Lo Weight, Cal Tolerance and Sensitivity
parameter settings.

On the Web page, left click on the Do Cal Lo button