On the HI 6300 series instrument display, select the calibration menu and use the UP or
DOWN button to select the C2 (eCAL) submenu, then select the Cal Tolerance (Calibration
Tolerance) menu item. The current value will be displayed, press the ENTER or the RIGHT
button if this value needs to be modified.

Do C2 (eCAL) Command

When this command is selected the C2 eCAL calibration process is run using the Cal
Tolerance, Gravity Correction, Ref Weight, and Sensitivity parameter settings.

NOTE

Ensure the Sensitivity, Gravity and Reference weight are all set prior to running C2 cal
command.

On the HI 6300 series instrument display, select the calibration menu and use the UP or
DOWN button to select the C2 (eCAL) submenu, then select C2 Cal. The instrument will
perform the calculation.

On the Web page, left click on the Do C2 eCAL button

Gravity Parameter

Gravity Correction compensates for an object weighing less at the equator than at the North
or South Pole.

Before modifying this value, please read the C2 calibration procedure described in Chapter
6.

On the Web page, select the calibration menu and left click inside the Gravity text field and
enter the desired Gravity value (see table in Chapter 6).

On the HI 6300 series instrument display, select the calibration menu and use the UP or
DOWN button to select the DO C2 eCAL submenu, then select the Gravity menu item. The
current value will be displayed, press the ENTER or the RIGHT button if this value needs
to be modified.