Calibration menu, C2 (ecal) commands and parameters – Hardy HI 6300 Series User Manual

Page 55

Advertising
background image

Instrument Configuration

47

Calibration Menu

C2 eCAL

o Sensitivity
o Gravity
o Ref Wt
o Do C2 Calibration
o Cal Tol
o Num Dev

Hard Cal

o Cal Tol
o Sensitivity
o Cal Lo Wt
o Do Cal Lo
o Span Wt
o Do Cal Hi

Cal Date

C2 (eCal) Commands and Parameters

Cal Tolerance Parameter

The value you enter for Calibration Tolerance sets the amount of deviation to allow during
the calibration process. This value must be greater than or equal to the base motion value
and/or the Graduation Sizes.

The base motion value can be calculated by using the following formula:

Base Motion value = (Total Load Cell Capacity x 0.0003)

We recommend three graduation sizes.

RANGE: .000001 - 999999 (default 10)

On the Web page, enter the value in the text box provided.