Usb memory stick parameters – Hardy HI 6300 Series User Manual

Page 70

Advertising
background image

62 Chapter

5

No specific baud rate is specified for printer, the Baud Rate parameter is used to set the baud
rate for the printer connected to the HI6310, typically 9,600 or 19,200.

On the Web page, select the communication menu, and left click inside the Baud Rate text
field and enter the Baud Rate.

On the HI6310 series instrument display, select the communication menu and use the UP
or DOWN button to select the Printer submenu, then select the Baud Rate menu item. The
current Baud Rate value will be displayed, press the ENTER or the RIGHT button if the
value needs to be modified.

Parity Bit Parameter

The Parity Bit parameter has three options, EVEN, ODD, or OFF.

On the Web page, select the communication menu, and then select the Printer submenu,
and left click inside the Parity field and select between EVEN, ODD, or OFF using the drop
down menu..

On the HI6310 series instrument display, select the communication menu and use
the UP or DOWN button to select the Printer submenu, then select the Parity menu
item. The current Parity value will be displayed, press the ENTER or the RIGHT
button if the Parity needs to be modified.

USB Memory Stick Parameters

Available on the HI6310 only

The HI6310 provides an interface to an external USB memory stick that allows HI 6310
parameters to be saved, restored or transfered to another instrument.

Restore Command

This command replaces the existing HI6310 parameters by the parameters stored on a USB
memory stick.

On the Web page, go the communication menu and left click on the Restore button.

On the HI6310 series instrument display, select the Communication menu and use the
UP or DOWN button to select the USB submenu, then select the Restore menu item then
press the ENTER button.

Advertising