Network Configuration 33

Modbus-RTU (over RS-485)

Step 1. Setup a slave address assigned to the HI 6300 series instrument from the

communications menu and the Modbus-RTU submenu.. Use a unique address
between 1 and 247

Step 2. Set the Baud Rate to the same as the master device, typically 9600 or 19,200

Step 3. Set the Parity Bit parameter to match that of the Modbus-RTU master (EVEN,

ODD, or OFF).

Step 4. Set the Termination to match the physical wiring of the serial port (ENABLE,

DISABLE, or LINK).

Wiring Pinout

The RS-485 serial port pinout is:

Modbus Setup

Slave Address may be set to any number in the range of 1-247.

Set Baud Rate parameter to match the settings of the Modbus master

Set Parity to match the settings of the Modbus master

Set Termination to Enable, Disable or Link to match the physical wiring.

NOTE

For a full list of I/O Communications parameters, see Appendix A.

Modbus Functions

The Modbus functions allowed in the HI 6300 are:

Function 3: Modbus Read Holding Registers

Function 4: Modbus Read Input Registers

Function 6: Modbus Write Single Register

Function 16 (0x10): Modbus Write Multiple Registers.

Modbus Registers

This instrument has 64 16-bit MODBUS input registers and 64 16-bit output registers. The
data in these registers is in "big endian" format, meaning that the most significant byte
comes first.

The first 6 registers (0 – 5) are used for the “command interface” in the unit.

Registers 6 & 7 will contain the Net weight in 32 bit floating point format.