On the HI 6300 series instrument display, select the calibration menu and use the UP or
DOWN button to select the C2 eCAL submenu, then select the Sensitivity menu item. The
list of sensitivity values supported will be displayed and the current value will be
highlighted, if the sensitivity needs to be modified use the UP and DOWN buttons to select
a new value and press the ENTER button.

Hard Cal Commands and Parameters

These commands and parameters are used when doing a Hard Cal or traditional calibration
of the instrument.

Cal Lo Weight Parameter

The value you enter for Cal Lo Weight sets the lower weight value to be used when
calibrating the system. The Cal Lo weight value, is normally zero (no weight on the scale),
but can be set to any known weight on the scale. With nothing (zero) on the scale, the Cal
Lo Weight is 0.00. With 5 lbs on the scale, the Cal Lo Weight is 5.00 lbs.

On the Web page, enter the value in the text box provided.

On the HI 6300 series instrument display, select the calibration menu and use the UP or
DOWN button to select the Hard Cal submenu, then select the Cal Lo Weight menu item.
The current value will be displayed, press the ENTER or the RIGHT button if this value
needs to be modified.