Span weight parameter – Hardy HI 6300 Series User Manual

Page 60

Advertising
background image

52 Chapter

5

On the HI 6300 series instrument display, select the calibration menu and use the

UP or DOWN button to select the Hard Cal submenu, then select the Do Cal Lo menu
item then press the ENTER button.

Do Cal Hi Command

Before executing this Do Cal Hi command verify that the Span Weight value is
correct. After setting/verifying the Span Weight value executing this command will
run the Hard Calibration process using the Cal Tolerance, Span Weight, and
Sensitivity parameter settings.

On the Web page, left click on the Do Cal High button as shown.

On the HI 6300 series instrument display, select the calibration menu and use the
UP or DOWN button to select the Hard Cal submenu, then select the Do Cal Hi menu
item then press the ENTER button.

Span Weight Parameter

The Span Weight value is the weight of the object that is being placed on the scale to
set the “High” calibration point with respect to the “Low” calibration point set using
the Cal Lo Weight parameter.

On the Web page, select the calibration menu and left click inside the Span Weight
text field and enter the load cell sensitivity value.