Front panel (optional), Load cell excitation, C2 calibration input – Hardy HI 6300 Series User Manual

Page 14: Cable lengths, Network connectivity, Environmental requirements, Operating temperature range, Temperature coefficient, Storage temperature range, 40 to 85º c (-40º to 185º f)

Advertising
background image

6 Chapter

2

Front Panel (optional)

Monochrome 480 x 272 LCD display with backlight

Five tactile keys for menu item selection

Load Cell Excitation

5 VDC +/- 1.15 VDC maximum

Isolation from digital section 1000 VDC minimum

C2 Calibration Input

Isolation from digital section 1000 VDC minimum

Cable Lengths

250 feet maximum of C2 authorized cable (Maximum of 4 load sensors) with IT

Junction box

Network Connectivity

Ethernet TCP/IP

Ethernet UDP

Modbus-RTU over RS485

HI 6310 Only
Analog 4-20 mA

Printer port

USB port

Environmental Requirements

Operating Temperature Range

0º to 60º C (32º to 140º F)

Temperature Coefficient

Less than 0.005% of full scale per degree C for Cal-LO and Cal-HI reference points

Storage Temperature Range

-40 to 85º C (-40º to 185º F)

Environmental

Intended for Building In, indoor use only at ambient temperatures between 10ºC to

60º C (14º to 140º F) with a pollution degree of 2.

Humidity Range

0-90% (non-condensing)

Maximum Altitude for Installation

2000 Meters (6,562 Feet)

Advertising