Parameters – Hardy HI 6300 Series User Manual

Page 150

Advertising
background image

140 Chapter

9

Parameters

Parameters is the first listed hyper-
link at the base of the Diagnostics
page. Click that link to display the
Parameters page. Note the scroll bar
on the right of the list.
The steps below explain how you
can duplicate the configuration of
one HI 6300 to use in configuring
another HI 6300 for sending a copy
of your parameters to
[email protected]
To copy the parameters:
Step 1. Right click in the

parameter list.

Step 2. Click Select All.

Step 3. Right click again.

Step 4. Click Copy.

To paste the parameter
configuration into the instrument you want to configure, or into a text file for storage:

Step 1. Enter the IP address of that instrument into a Web browser to display its Home

page.

Step 2. Select Operation, then Diagnostics, then Parameters, as described above.

Step 3. Right click in the Parameter list.

Step 4. Click Select All.

Step 5. Right click again in the Parameter List.

Step 6. Click on Paste to replace the existing or default parameter settings with the

parameters settings of the instrument you copied. Only the past setting from a text
file will work---you can not paste from a word processing program.

Step 7. Click the Save button. You now have an exact duplicate of the pre-configured

instrument.

If you need to make any modifications to the parameter settings go to Chapter 5,
Configuration, for more instructions.