The Termination Parameter has three options, DISABLE, ENABLE and LINK.

Disable: The Modbus-RTU signal is not terminated

Enable: The Modbus-RTU signal is terminated

Link: Manual Termination.

On the Web page. select the Termination drop down and select the termination condition.
Link means that the termination is set on or off depending on whether pin 2 of the connector
is low (on) or high (off). There is an internal pull up on pin 2 that will set the termination
off if nothing is connected to it.

Select Save Parameter to save the entry.

Click Home to return to the home page.

On the HI 6300 series instrument display, select the communication menu and use
the UP or DOWN button to select the Modbus-RTU submenu, then select the Slave
Address menu item. The current Termination value will be displayed, press the
ENTER or the RIGHT button to change the parameter if the Termination needs to be
modified.

Printer Command Parameters

Available on the HI6310 only

RS-232 printer configuration.

The HI 6310 series can transmit Gross, Net, Count, and Tare values to a serial printer over
RS-232. The printer Baud rate is user selectable at 600, 1200, 2400, 4800, 9600, or 19200.