Specifications

5

Chapter 2

• • • • • •

Specifications

Chapter 2 provides specifications for HI 6300 series instruments. The specifications
listed are designed to assist in the installation, operation and troubleshooting of your
instrument. All service personnel should be familiar with this section before installing
or repairing the instrument.

Basic Specifications

General

Number of Channels

1 Channel

Update Rate

55 Updates per Second

Resolution

Internal 1:262,144

WAVERSAVER

User Selectable

OFF

7.50 Hz

3.50 Hz

1.00 Hz (default)

Averages

1 to 250 User-selectable in Single Increments

Power

12-24 VDC (Class 2 Source)

Input

Up to four 350-ohm full Wheatstone bridge, strain gauge load sensor/cells (5 volt

excitation) can be connected to the weigh scale input

Non-linearity

0.0015% of full scale

Common Mode Rejection

110dB at or below 60 Hz

Common Mode Voltage Range

2.5 VDC maximum (with respect to earth ground)