Instruction Manual

43

Calibrating the Meter with External Source (Cal)

To scale the meter without a signal source

refer to Scaling the Meter (SCALE) page 39.

The meter can be calibrated to display the process variable in engineer-
ing units by applying the appropriate input signal and following the cali-
bration procedure.
The use of a calibrated signal source is strongly recommended to cali-
brate the meter.

CAL

Prog

Inp 1

CAL

Inp 1

I n p 1

Inp 2

CAL

Inp 2

I n p 2

dis 1

Inp 1

00

0

.000

d i s 1

dis 2

Inp 2

1

00.000

d i s 2

T he display

flashes while

sensing the

inputs

Warm up the meter for at least 15 minutes

before performing calibration to ensure

specified accuracy.