4 electrical data, 5 consumption data, Electrical data – YSI P 700 IQ Analyzer User Manual

Page 104: Consumption data, 4 electrical data 7.5 consumption data

Advertising
background image

104

ba76133e01

09/2013

P 700 IQ

7.4 Electrical data

7.5 Consumption data

The consumption of chemicals depends on the specified intervals and
on the selected measuring range.

Typical consumption

values

Input voltage

Depending on the selected variant:

115 V AC, 60 Hz

Fuse protection of the
building

16 A with ground fault circuit inter-
rupter

Power consumption

35 - 1100 W depending on the option

• Basic consumption

35 W

• Filtration pump

40 W

• Climate kit

200 W

• Heat tracing of the

intake tube

20 W/m (max. 20 m allowed)

• Heat tracing of the

return tube

20 W/m (max. 20 m allowed)

Power supply

Nominal voltage:
115 V AC ± 10 % / 60 Hz

Power supply:
3-pin, N/L /PE

Cable cross-section of power supply:
Europe: 1.5 ... 4.0 mm

2

USA: AWG 14 ... 12

Fuse protection on the operator side:
16 A maximum

Solution

Sufficient for

Reagent solution RE 2.5

4 months (measuring range A, mea-
suring interval 5 minutes)

4 months (measuring range B, mea-
suring interval 10 minutes)

Cleaning solution CL 1.0

4 months (daily cleaning)

Standard solution ST

4 months (daily calibration)