YSI P 700 IQ Analyzer User Manual

Page 106

Advertising
background image

106

ba76133e01

09/2013

P 700 IQ

EAM3Bx

Meas. range exceeded or undercut

* Check process

EC13Bx

Automatic calibration error: Calibration standard could not be deter-
mined or is not suitable for current measurement range. System
stopped!

* Check the current measurement range

* Check the used calibration standard

EC13Bx

Sensor could not be calibrated,
Sensor blocked for measurement

* Check calibration conditions and calibration standard

* View calibration history

* Service sensor immediately
(see operating manual)

EI13Bx

Operational voltage too low

* Check installation and cable lengths, Follow installation instructions

* Power supply module(s) overloaded, add power supply module(s)

* Check terminal and module connections

EI23Bx

Operational voltage too low, no operation possible

* Check installation and cable lengths, Follow installation instructions

* Power supply module(s) overloaded, add power supply module(s)

* Check terminal and module connections

ES13Bx

Component hardware defective

* Contact service

IC13Bx

Factory calibration has been activated. Make sure the P700IQ-PO4
operates correctly.

IC13Bx

Sensor has been successfully calibrated

* For calibration data, see calibration history

IC23Bx

Last valid user calibration has been activated. Make sure the
P700IQ-PO4 operates correctly.

IC33Bx

Invalid user calibration has been replaced by last valid user calibra-
tion. Caution! Wrong measured values possible. Carry out a new
successful calibration to make sure the P700IQ-PO4 operates cor-
rectly.

IC43Bx

The invalid user calibration was replaced by the factory calibration.
Caution! Wrong measured values possible. Check whether the
P700IQ-PO4 operates correctly or carry out a new successful user
calibration.

Message code

Message text