5 connecting the lines – YSI P 700 IQ Analyzer User Manual

Page 38

Advertising
background image

38

ba76133e01

09/2013

P 700 IQ

3.3.5 Connecting the lines

Fig. 3-16,  38 shows the housing with the completely connected lines
including the accessories and all options:

Fig. 3-16: Completely connected lines

Proceed as follows to connect the individual lines:

1

Connect the intake tube (2) to the filtration pump.

2

Insert the outlet of the overflow vessel and the outlet of the pho-
tometer unit into the sleeve tube of the return tube.

3

Connect the IQ S

ENSOR

N

ET

cable to the MIQ/WCA 232. See

section 3.3.5.1 Connecting the IQ S

ENSOR

N

ET

cable,  39.

4

Connect the power line (1) and, if necessary, the heat tracing
lines (3 and 4) to the power supply box. See section 3.3.5.2 Con-
necting the power line and heat tracing lines,  40.

5

Tighten all cable glands after the connections have been made.

1

2

3

4

6

5

7

1 Power line
2 Intake tube
3 Heat tracing of the suction line (option)
4 Heat tracing of the return tube (option)

5 IQ S

ENSOR

N

ET

cable

6 Outlet of the overflow vessel
7 Outlet of the photometer unit

The liquid from the return tube must be able to drain off
freely (downward slope).