YSI P 700 IQ Analyzer User Manual
Page 108
108
ba76133e01
09/2013
P 700 IQ
Status info,
general structure
The bits 0 - 15 are reserved for general information.
The bits 16 - 21 are reserved for internal service information.
You obtain the status info:
• via a manual query in the Einstellungen/Setup/Service/List of all
components menu (see system operating manual)
• by an automated query
• from a superordinate process control (e. g. when connected to
the Profibus)
• from the IQ Data Server (see IQ S
ENSOR
N
ET
Software Pack
operating manual)
The evaluation of the status info, e.g. in the case of an automated
query, has to be made individually for each bit.
Status info
P 700 IQ
0 1 2 3 4 5 6 7
8 9 10 11 12 13 14 15
1 0 0 0 0 0 0 0
0 0 0 0 0 0 0 0
(general)
0 0 0 0 0 0 0 0
0 0 0 0 0 0 0 0
(internal)
16 17 18 19 20 21 22 23
24 25 26 27 28 29 30 31
Status bit
Explanation
Bit 0
Component hardware defective
Bit 1
Liquid tank xxx almost depleted!
Bit 2
Air bubbles in the analyzer
Bit 3-31
-