YSI P 700 IQ Analyzer User Manual

Page 52

Advertising
background image

52

ba76133e01

09/2013

P 700 IQ

4

Then navigate to the setting menu of the P 700 IQ. The exact pro-
cedure is given in the relevant IQ S

ENSOR

N

ET

system operating

manual.

5

Adjust the setting values as necessary.

Setting table

Tab. 4-2,  53 shows the setting menus with the possible values to be
set. Default values are marked in bold.

Setting menu

Possible values

Description

Measuring mode

PO4-P

PO4

The measured parameter is displayed in
the selected citation form.

Measuring range

Display of measured values

Measuring mode PO4-P

A: 0.05 .. 15.00 mg/L
B: 1 ... 50 mg/L

A: 0.05 ... 15.00 ppm
B: 1 ... 50 ppm

For each measuring mode (PO4 / PO4-
P) and each displayed unit (mg/L, ppm),
a small measuring range (A) and a large
measuring range (B) can be selected.

The measuring range can only be
selected when the P 700 IQ is stopped
(calibration and service menu).

Measuring mode PO4

A: 0,15 .. 46,00 mg/L
B: 3 ... 153 mg/L

A: 0.15 ... 46.00 ppm
B: 3 ... 153 ppm

Offset correction

A: - 0.50…0.00…+ 0.50
mg/L

B: - 5…0…+ 5 mg/L

A: Setting range for:
Measuring mode PO4-P and

Measuring range 0.05 .. 15.00 mg/L
B: Setting range for:
Measuring mode PO4-P and

Measuring range 1 ... 50 mg/L
For other measuring modes and measur-
ing ranges, the value ranges are
adjusted.

Meas. interval

5, 10, 15, 20, 30, 45, 60
min

2, 4, 6, 8 , 12 h

Measuring interval

(Smallest measuring interval for measur-
ing range B: 10 min.)

Signal smoothing

On

Off

Signal filter

A signal filter in the sensor reduces the
limits of variation of the measured value.

The signal filter is essentially character-
ized by the response time t90.

Response time t90

100, 200, 300, 400 s

Response time of the signal filter (in sec-
onds).

This is the time after which 90 % of a
signal change is displayed.