YSI P 700 IQ Analyzer User Manual
Page 56
56
ba76133e01
09/2013
P 700 IQ
lowing standards can be used for manual 1-point calibration: ST 0.0,
ST 1.6, ST 8.0, ST 20.0.
Proceed as follows to carry out a 1-point calibration procedure manu-
ally:
1
For conveniently working on site:
Dock an IQ S
ENSOR
N
ET
terminal onto an MIQ module in the vicin-
ity of the analyzer, or onto the MIQ/WCA 232
module in the ana-
lyzer.
Fig. 4-3:
IQ S
ENSOR
N
ET
- terminal
Stopping the running
operation
2
Using <>, select the P 700 IQ in the measured value display.
Fig. 4-4: Measured value display
3
Open the calibration and service menu with <C>.
The maintenance condition is activated.
4
Confirm the maintenance condition with <OK>.
The calibration and service menu is displayed.
A manual 1-point calibration with the same calibration stan-
dard used for automatic calibration can be carried out with-
out opening the enclosure.