YSI P 700 IQ Analyzer User Manual

Page 56

Advertising
background image

56

ba76133e01

09/2013

P 700 IQ

lowing standards can be used for manual 1-point calibration: ST 0.0,
ST 1.6, ST 8.0, ST 20.0.

Proceed as follows to carry out a 1-point calibration procedure manu-
ally:

1

For conveniently working on site:
Dock an IQ S

ENSOR

N

ET

terminal onto an MIQ module in the vicin-

ity of the analyzer, or onto the MIQ/WCA 232

module in the ana-

lyzer.

Fig. 4-3:

IQ S

ENSOR

N

ET

- terminal

Stopping the running

operation

2

Using <>, select the P 700 IQ in the measured value display.

Fig. 4-4: Measured value display

3

Open the calibration and service menu with <C>.
The maintenance condition is activated.

4

Confirm the maintenance condition with <OK>.
The calibration and service menu is displayed.

A manual 1-point calibration with the same calibration stan-
dard used for automatic calibration can be carried out with-
out opening the enclosure.

Advertising