Carry out the steps 15…20 in section 5.3.4 Maintenance and
cleaning work on the open P 700 IQ,  73.

5.4.2 Changing the filter mats

The ventilation grids are under the enclosure lid and on the underside
of the enclosure. The ventilator is behind a protective grating. The filter
mat can thus be exchanged riskless even during operation.

Maintenance interval

Depending on the contamination

Upper ventilation grid

1

Remove both front screws on the enclosure lid.

2

Open the enclosure lid upward and fix it with the bearer. See
Fig. 5-6,  78.

Fig. 5-6: Opening the enclosure lid

1 Enclosure lid
2 Bearer
3 Fixing screw

3

Open the ventilation grid with the aid of a screwdriver by levering
it off and exchange the filter mat. See Fig. 5-7,  78.

Fig. 5-7: Exchanging the upper filter mat