YSI P 700 IQ Analyzer User Manual

Page 96

Advertising
background image

96

ba76133e01

09/2013

P 700 IQ

Display of OFL

Display of ERROR

Interferences
e.g. in sample matrix

Select suitable measuring loca-
tion

A valve does not operate

Perform a Hydraulic check

Contact the service department

Photometer unit draws air (e.g.
lack of sample in the overflow
vessel)

Make sure that there is enough
sample in the overflow ves-
sel,e.g.

Service the filter unit

Set the pump output to 10 - 100
ml / min

Cause

Remedy

Cause

Remedy

Measuring range exceeded

Select different measuring range

Calibration error

Carry out calibration

Measuring cell dirty

Carry out cleaning procedure

Cause

Remedy

P700IQ-PO4

not correctly con-

nected

Check the assignment of the ter-
minal connections

Load at a power supply module
too high

Install an additional power
supply module (MIQ/PS) in the
vicinity of the sensor