YSI P 700 IQ Analyzer User Manual

Page 75

Advertising
background image

ba76133e01

09/2013

75

P 700 IQ

If necessary,

Empty the system

6

Carry out the Empty the system function for the following mainte-
nance activities (otherwise, continue with step 12):
• Change the valve tubes (T SET)

Changing the pump tube of the peristaltic pump

7

Follow the instructions on the display.

8

Place all intake tubes into deionized water.

9

Start the rinsing procedure.

10

After the rinsing, leave the intake tubes in air.

11

Carry out the Empty the system function once again. After the
emptying is finished, the calibration and service menu is dis-
played again.

12

Carry out the required maintenance activities, such as:
• Change the liquid containers

(see section 5.4.1,  76)

• Change the valve tubes (T SET)

(see section 5.4.3 Changing the tube set (T SET),  79)

Changing the pump tube of the peristaltic pump

(see section 5.4.4,  83)

• Clean the overflow vessel

(see section 5.4.7 Cleaning the overflow vessel,  93)

• Change the filter membrane

(see section 5.4.6 Exchanging the filter membrane,  92)

• Clean the filter membrane

(see section 5.4.5 Cleaning the filter membrane,  88)

• Perform Hydraulic check

(see section 5.4.8 Performing a Hydraulic check,  94)

13

If the Empty the system function was carried out:
Carry out the Fill the system function after the maintenance activ-
ities are finished. All valve tubes are automatically filled with the
connected solutions and sample liquid.

14

Optional (as necessary):
Carry out the Purge manually function to rinse the measuring
cycle with cleaning solution.

Restarting the

measuring operation

15

Carry out the Start measuring function. Measurement is started
and the measured value is displayed in the measured value dis-
play after approx. 5…7 minutes.

16

Carry out the Exit SERVICE function.
or
Exit the calibration and service menu with <ESC>.

17

Confirm the note with <OK>. The measured value display
appears. The measured value flashes.