P 700 IQ

Start measuring

Procedure to start a measurement.

The functionality of the photometer unit is
unblocked. A first measurement is started.

After the first measurement, further clean-
ing/calibration/measuring cycles will follow
according to the specified intervals.

Used in the following situations:
• Measuring operation is restarted (e.g. after

maintenance activities)

• To compete the initial commissioning

Calibrate manually

Procedure to carry out a manual calibration
(1-point-cal. or 2-point-cal.)

Used in the following situations:
• Measured values are implausible

• Calibration result implausible

• Maintenance activities on liquids, tubes or

measuring cycle were carried out

Fill the system

Procedure to fill all tubes with liquids and sam-
ple.

Used in the following situations:
• The tubes are not completely (to the valve)

filled with liquid, e.g. after exchanging the
liquid containers

• Air is in the tube

System maintenan
ce

Procedure to exchange containers (liquids):

Sets the counter for the selected container to
100% and fills the relevant tubes. Selection:

Replace reagent

Replace standard

Replace cleaning solution

Replace all
(reagent, standard and cleaning solution)

Return

Hydraulic check

After the Hydraulic check function was started

• all 5 valves are activated one after the

other

• the motor of the pump is tested in 3 condi-

tions (slowly forwards, quickly forwards,
backwards)

Continue . . .