YSI P 700 IQ Analyzer User Manual
Page 59
Advertising
ba76133e01
09/2013
59
P 700 IQ
Fig. 4-6:
IQ S
ENSOR
N
ET
- terminal
Stopping the running
operation
4
Using <>, select the P 700 IQ in the measured value display.
5
Open the calibration and service menu with <C>.
The maintenance condition is activated.
6
Confirm the maintenance condition with <OK>.
The calibration and service menu is displayed.
Fig. 4-7: Calibration and service menu
7
Carry out the Stop function to stop the running operation.
8
Option:
Carry out the Purge manually function to rinse the measuring
cycle with cleaning solution.
Carrying out calibration
9
Carry out the Calibrate manually / 2-point-cal. function.
10
Follow the instructions on the display.
The order of the calibration standards is not important.