YSI P 700 IQ Analyzer User Manual

Page 80

Advertising
background image

80

ba76133e01

09/2013

P 700 IQ

Tab. 5-5: Tubes of the T-SET (order no 821992B)

for photometer unit with plastic connectors

Maintenance interval

12 months

Preparations

1

Carry out the steps 1…11 in section 5.3.4 Maintenance and
cleaning work on the open P 700 IQ,  73.

2

Disconnect the plug from the P700IQ-PO4
(the heat tracing remains operating).
or
Switch off the analyzer with the power switch.

3

Screw off the empty photometer unit.

4

Carry out the mounting work on a lab table.

5

Remove the plastic cover from the photometer unit.

140

Cell (plastic connection) -
V8 C

1 nut
1 reinforced

120
+

Sample tube 400 mm -
sample

1 luer

400 V11 NC

-

sample tube 120 mm

1 reinforced
1 luer

400

Green (reagent)

VK5 NO -
reagent

1 reinforced
1 luer

400

Orange (calibration
standard)

VK6 NO - calibration stan-
dard

1 reinforced
1 luer

400

Blue (cleaning)

VK6 NC - cleaning solution 1 reinforced

1 luer

500

V8 NO - waste container

1 reinforced

Tube length
(mm)

Color coding

Connection of

Tube ends