YSI P 700 IQ Analyzer User Manual
Page 103
ba76133e01
09/2013
103
P 700 IQ
Instrument safety
Test certificates
CE, cETLus
Light source for photome-
ter
LED, 420 nm (violet)
Filtration unit
Membrane surface: 1000 cm
2
Cut-off limit: < 0.45 µm
Max. operating temperature: 45 °C
(113 °F)
pH value: 2…11,5
Housing: PVC
Seals: NBR
Membrane: PVDF membrane, porous,
glued to polyester fleece on both sides
Sleeve tube:
PVC-reinforced PCV tube, transpar-
ent, 24 x 3 mm
Intake tube: PE 4 x 1 mm
Tube length (inlet and outlet tube):
Max. 20 m
Overflow vessel
Material: PVC
Required sample quantity:
10 ... 100 ml/min
For correct measurements, enough
sample must always be available in the
overflow vessel.
Applicable norms
EN 61010-1
UL 61010-1
CAN/CSA C22.2#61010-1
Electromagnetic compati-
bility
EN 61326-1, EN 61326-2-3,
FCC 47 CFR Part 15
Type of protection (enclo-
sure)
IP 54 (EN 60529)
Protective class
I
Overvoltage category
II