YSI P 700 IQ Analyzer User Manual
Page 74
74
ba76133e01
09/2013
P 700 IQ
Opening the enclosure
1
Open the enclosure of the P 700 IQ.
Selecting the P 700 IQ
2
Using <>, select the P 700 IQ in the measured value display.
Fig. 5-3: Measured value display
Stopping the running
operation
3
Using <C>, switch to the calibration and service menu.
The maintenance condition is activated.
4
Confirm the maintenance condition with <OK>.
The calibration and service menu is displayed.
Fig. 5-4: Calibration and service menu
5
Carry out the Stop function to stop the running operation.
Important for maintenance activities in the open:
If the ambient temperatures are under 0 °C, make sure that
the liquids do not freeze while the maintenance activities
are carried out. Leave the enclosure open for as short a
time as possible.