YSI P 700 IQ Analyzer User Manual

Page 74

Advertising
background image

74

ba76133e01

09/2013

P 700 IQ

Opening the enclosure

1

Open the enclosure of the P 700 IQ.

Selecting the P 700 IQ

2

Using <>, select the P 700 IQ in the measured value display.

Fig. 5-3: Measured value display

Stopping the running

operation

3

Using <C>, switch to the calibration and service menu.
The maintenance condition is activated.

4

Confirm the maintenance condition with <OK>.
The calibration and service menu is displayed.

Fig. 5-4: Calibration and service menu

5

Carry out the Stop function to stop the running operation.

Important for maintenance activities in the open:
If the ambient temperatures are under 0 °C, make sure that
the liquids do not freeze while the maintenance activities
are carried out. Leave the enclosure open for as short a
time as possible.