If necessary, disconnect the terminal from the P 700 IQ.

19

Close the enclosure of the P 700 IQ.

20

Wait until a measured value is shown in the measured value dis-
play (approx. 5…7 minutes).

21

Switch off the maintenance condition. See IQ S

ENSOR

N

ET

system operating manual.

5.4 Carrying out the maintenance and

cleaning activities

5.4.1 Changing the liquid containers

The P 700 IQ has an extra counter for each liquid container (from soft-
ware version 2.13). The counter counts the consumption of the follow-
ing procedures as soon as the function was started:

• Measuring

• Calibrate

• Cleaning

Other procedures that consume liquids are not counted (e.g. rinsing,
filling).

The counter displays the remaining quantity as a percentage.

If the remaining quantity of a liquid is only approx. 10% of the capacity,
a log book message is automatically issued.

Replace the container before the liquid is used up.

Maintenance interval

Approx. 4 months,
depending on the frequency of measurement, cleaning, calibration.

The current status of the counters for the liquid containers
can be output as log book messages (sensor settings /Sta-
tus (P700IQ-PO4)
/Transmit to log book).

The filling level can only be displayed correctly when the
new container is full (filling level 100 %).