5 reactivating the last valid calibration – YSI P 700 IQ Analyzer User Manual

Page 62

Advertising
background image

62

ba76133e01

09/2013

P 700 IQ

t (offset)

• Calibration evaluation:

• +: Successful calibration.

The new calibration data are taken over for measurement.

• -: Calibration unsuccessful.

Sensor blocked for measurement.

4.2.5 Reactivating the last valid calibration

If a manual calibration procedure is unsuccessful, the measuring oper-
ation can only be resumed after the manual calibration was repeated or
a valid calibration was reactivated.

If an automatic calibration procedure is unsuccessful, the analyzer
stops. To be able to measure again, you have to carry out a valid cali-
bration procedure in the calibration and service menu, or to reactivate
a valid calibration.

Thus you can immediately continue to measure if a calibration failed or
you suspect that the calibration conditions were not optimally met.

1

Using <>, select the P 700 IQ in the measured value display.

2

Using <C>, switch to the calibration and service menu.
The maintenance condition is activated.

3

Confirm the maintenance condition with <OK>.
The calibration and service menu is displayed.

4

Carry out the Start measuring function.
A menu to select the last valid calibration or the factory calibration
is displayed.

5

Select the last valid calibration or the factory calibration.

6

Carry out the Exit SERVICE function.
or
Exit the calibration and service menu with <ESC>.

7

Confirm the note with <OK>. The measured value display
appears. The measured value flashes.

8

If necessary, disconnect the terminal from the MIQ/WCA 232.

9

Close the enclosure of the P 700 IQ.

Reactivating old calibration data is a temporary measure.
Take into consideration that the sensor my provide wrong
measured values. Ensure the correct functioning of the
sensor by checking and/or recalibrating it.

Advertising