YSI P 700 IQ Analyzer User Manual

Page 50

Advertising
background image

50

ba76133e01

09/2013

P 700 IQ

Fig. 4-1: Calibration and service menu

4

If the system is not yet filled:
Carry out the Fill the system function.
All tubes of the P700IQ-PO4 are automatically filled with the con-
nected solutions and sample liquid.

5

Carry out the Start measuring function.
The measured value determination takes approx. 5…7 minutes.

6

Follow the instructions on the display.

7

Carry out the Exit SERVICE function.
or
Exit the calibration and service menu with <ESC>.
A note to end the maintenance condition is displayed.

8

Confirm the note with <OK>.
The measured value display appears. While no valid measured
value is available, the display shows bars «- - - -».

9

Wait until a measured value is shown in the measured value dis-
play (approx. 5…7 minutes).

10

Switch off the maintenance condition.
See IQ S

ENSOR

N

ET

system operating manual.

During measuring operation, the measurements take place at the spec-
ified intervals.

If the measurement is delayed because a planned cleaning
or calibration procedure is carried out beforehand, an info
message is displayed.