YSI P 700 IQ Analyzer User Manual
Page 50
50
ba76133e01
09/2013
P 700 IQ
Fig. 4-1: Calibration and service menu
4
If the system is not yet filled:
Carry out the Fill the system function.
All tubes of the P700IQ-PO4 are automatically filled with the con-
nected solutions and sample liquid.
5
Carry out the Start measuring function.
The measured value determination takes approx. 5…7 minutes.
6
Follow the instructions on the display.
7
Carry out the Exit SERVICE function.
or
Exit the calibration and service menu with <ESC>.
A note to end the maintenance condition is displayed.
8
Confirm the note with <OK>.
The measured value display appears. While no valid measured
value is available, the display shows bars «- - - -».
9
Wait until a measured value is shown in the measured value dis-
play (approx. 5…7 minutes).
10
Switch off the maintenance condition.
See IQ S
ENSOR
N
ET
system operating manual.
During measuring operation, the measurements take place at the spec-
ified intervals.
If the measurement is delayed because a planned cleaning
or calibration procedure is carried out beforehand, an info
message is displayed.