YSI P 700 IQ Analyzer User Manual

Page 60

Advertising
background image

60

ba76133e01

09/2013

P 700 IQ

11

Calibrate with calibration standard 1.

11.1 Connect calibration standard 1.

11.2 Select calibration standard 1.

11.3 Start the calibration procedure with <OK>.

12

Connect calibration standard 2 (see chapter 3.3.5.3 Connecting
the liquid containers,  42)
.

13

Calibrate with calibration standard 2.

13.1 Connect calibration standard 2.

13.2 Select calibration standard 2.

13.3 Continue the calibration procedure with <CONTINUE>.

The calibration result is displayed after calibration standard 2 has been
measured (Calibration successful!).

Restarting the

measuring operation

14

If necessary, reconnect the calibration standard for automatic
calibration.

15

Carry out the Start measuring function.
Measurement is started and the measured value is displayed in
the measured value display after approx. 5…7 minutes.

16

Carry out the Exit SERVICE function.
or
Exit the calibration and service menu with <ESC>.

17

Confirm the note with <OK>.
The measured value display appears. The measured value
flashes.

18

If necessary, disconnect the terminal from the MIQ/WCA 232.

19

Close the enclosure of the P 700 IQ.

You have the following possibilities if the calibration failed:
• Repeat the calibration procedure

(make sure that the correct calibration standard is con-
nected and selected for calibration).

• Use the last valid calibration

(see section 4.2.5 Reactivating the last valid calibration,

62 )

• Use the factory calibration

(see section 4.2.5 Reactivating the last valid calibration,

62)