4.2 Calibration

4.2.1 Overview

Why calibrate?

During operation, components of the P 700 IQ can age or become dirty
and thus change their characteristics, e.g.:

• Flow-through cell

• LED

• Photo diodes

• Color reagent

Regular automatic or manual calibration procedures help you to recog-
nize any changes of the measuring characteristics of the P 700 IQ.

When to calibrate?

Calibrate at regular intervals.

1-point calibration

The offset of the characteristic curve can be adjusted with a regular 1-
point calibration.

The 1-point calibration can be carried out automatically at regular inter-
vals. An additional 1-point calibration procedure can be started in the
calibration and service menu at any time. See section 4.2.2,  55.

The Autom.cleaning, Autom.calibration and Meas. interval
settings may result in the overlapping of the carrying out of
different functions.
In this case, the functions are carried out according to pri-
ority.

Function

Priority

Duration
(min)

Measuring
range

Autom.cleaning

1

5

A, B

Autom.calibration

2

6

9

A

B

Measurement

3

5

7

A

B