YSI P 700 IQ Analyzer User Manual

Page 72

Advertising
background image

72

ba76133e01

09/2013

P 700 IQ

Fig. 5-1: Measured value display

Stopping the running

operation

2

Using <C>, switch to the calibration and service menu.
The maintenance condition is activated.

3

Confirm the maintenance condition with <OK>.
The calibration and service menu is displayed.

Fig. 5-2: Calibration and service menu

4

Carry out the Stop function to stop the running operation.

Carrying out

maintenance activities

5

Carry out the required maintenance activities, such as:
• Start the maintenance function in the calibration and service

menu (Purge manually, Calibrate manually / 1-point-cal.)
Follow the instructions on the display.
After the maintenance function is finished, the calibration and
service menu is displayed again.

• Changing the filter mats

(see section 5.4.2 Changing the filter mats,  78)