Measuring mode cannot

be set

IQ S

ENSOR

N

ET

cable towards

the P 700 IQ too long (voltage
drop too great)

Install an additional power
supply module (MIQ/PS) in the
vicinity of the sensor

Electrical connection between
power supply module and
P 700 IQ is interrupted

Check the cable connection
step-by-step starting with the
power supply module (MIQ/PS),
and replace defective cable sec-
tions

Check the contacts on the MIQ
modules (stack mounting).
Clean any dirty contacts. Care-
fully bend back contact springs
that have been pressed flat or
bent (pay attention to sufficient
spring tension)

External electrical power supply
disturbed

Check the power supply,

check the maximum load

Cause

Remedy

Cause

Remedy

P 700 IQ is not stopped

Stop the
P 700 IQ (see section 5.3.2,

69).