P 700 IQ

Tab. 4-2: Setting menu

Autom.cleaning

On

Off

Switches the automatic cleaning function
on or off

Interval

6, 12, 24, 48, 96

Interval in h

Ref.time hours

0…7…23

Defines the start time from which the
automatic cleaning procedures will take
place at the set interval

(default setting: 7:00 o'clock)

Ref.time minutes

0…59

Autom.calibration

On

Off

Switches the automatic calibration func-
tion on or off

Interval

6, 12, 24, 48, 96

Interval in h

Ref.time hours

0…8…23

Defines the start time from which the
automatic calibration procedures will
take place at the set interval

(default setting: 8:00 o'clock)

Ref.time minutes

0…59

Status (P700IQ-PO4)

A log book message with the current
status is output.

The status indicates the current situation
of the P700IQ-PO4, e.g.
• Cleaning (ON/OFF)

• Calibrating (ON/OFF/WAIT CLEAN)

• Measuring (ON/OFF/WAIT

CLEAN/WAIT CAL)

• Filling levels of the liquid containers in

%

Meaning
• ON: Function active

• WAIT: Waiting for a cleaning cycle

(WAIT CLEAN) or calibration cycle
(WAIT CAL) to be performed at a
specified point of time.

Autostart

On

Off

Activate or switch off the automatic start
of the P 700 IQ after a power failure

Save and quit

The P 700 IQ stores all changed settings
and the display switches to the next
higher level.

Quit

The display switches to the next higher
level without storing the new settings.

Setting menu

Possible values

Description