Fixing the rail mount

bracket

4

Place the rail mount bracket in front of the rail in the required posi-
tion.

Fig. 3-9: Mounting the rail mount bracket on the rail

5

Attach the rail mount bracket to two suitable horizontal rail pipes
with the aid of the four clamps. Each clamp consists of 2 slot bars
(6), 2 long hexagon head screws M8x80 (7), 2 nuts (8) and 2
spring washers (9). Adjust the positions of the clamps to the rail
pipes. Note that the rail mount bracket must still stand on the
ground!

The weight of the analyzer is supported by the rail mount
bracket standing on the ground. The rail prevents the ana-
lyzer from tilting over.