Indexes, 1 explanation of the messages, 1 error messages – YSI P 700 IQ Analyzer User Manual

Page 105: Explanation of the messages 8.1.1 error messages

Advertising
background image

ba76133e01

09/2013

105

P 700 IQ

8. Indexes

8.1 Explanation of the messages

This chapter contains a list of all message codes and the related mes-
sage texts that can occur in the log book of the IQ S

ENSOR

N

ET

system

for the P 700 IQ analyzer.

The last three digits of the message code form the component code. It
identifies the component (active component) that caused the message:

Some error messages contain an internal error code, starting with "#".

8.1.1 Error messages

Information on the contents and structure of the log book

and the structure of the message code is given in the L

OG

BOOK

chapter of the IQ S

ENSOR

N

ET

system operating man-

ual.

Component code

Component

3Bx

P700IQ-PO4

55x

MIQ/WCA 232

Message code

Message text

EA13Bx

Meas. range exceeded or undercut

* Check process

* Select other meas. range

EA43Bx

Analyzer leakage

* Check the tube connections

EA53Bx

Physical measurement range exceeded

* Check the process

EA63Bx

Air bubbles in the analyzer

* Check the tube connections

* Check fuel levels and sample flow

* Check the process

Advertising