2 info messages, 2 status info, Status info – YSI P 700 IQ Analyzer User Manual

Page 107

Advertising
background image

ba76133e01

09/2013

107

P 700 IQ

8.1.2 Info messages

The analyzer does not generate any info messages.

8.2 Status info

The status info is a coded piece of information on the current status of
a sensor. Each sensor sends this status info to the controller. The
status info of sensors consists of 32 bits, each of which can have the
value 0 or 1.

IC53Bx

P700IQ-PO4 was successfully calibrated

* For calibration data, see calibration history

II13Bx

Language not available,
Default language German

* Contact service

IC63Bx

Liquid tank xxx almost depleted!

* Please refill to maintain proper functioning

EI155x

Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)

* Check terminal and module connections
* Defective components, replace components

EI255x

Operational voltage too low, no operation possible

* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
* Defective components,
replace components

II155x

Language not available,
Default language German
* Contact service

ES255x

Component hardware defective xxx

* Contact service

EIA55x

Communication between MIQ/WCA and P700IQ-PO4 disturbed

* Check cable connections
* Provide voltage supply to P700IQ-PO4
* Contact service

Message code

Message text

Advertising