3 installation on a wall, See section 3.3.2.3 installation on a wall,  33, Attention – YSI P 700 IQ Analyzer User Manual

Page 33

Advertising
background image

ba76133e01

09/2013

33

P 700 IQ

3.3.2.3 Installation on a wall

With the rails on its rear side, the analyzer is hooked into the in die
retaining hooks of the WM wall mounting set.

Proceed as follows to install the housing on a wall:

1

Drill eight holes as shown in the following figure:

Fig. 3-12: Drilling dimensions for mounting the WM wall mount

2

Screw tight the four retaining hooks of the wall mounting set.

3

Mount the housing by hooking the rails fixed on its rear side into
the retaining hooks. The fixing screws of the rail must be outside
the retaining hooks on both sides as shown in the following figure:

ATTENTION

Make sure that the wall is strong enough for the weight
of the analyzer and that the mounting material (screws,
plugs, etc.) is suitable for the wall type. If necessary, use
other screws and plugs than the ones provided.

540

408

30

Recommended height above the ground approx. 165 m

Advertising