3 installing the fm filter membrane module and the, M 1.5 attachment for filtration, See section 3.3.3,  35 – YSI P 700 IQ Analyzer User Manual

Page 35

Advertising
background image

ba76133e01

09/2013

35

P 700 IQ

3.3.3 Installing the FM filter membrane module and the

M 1.5 attachment for filtration

Installation instructions

Heed the following notes when installing the filter membrane module:

• Mount the filter membrane module so that the plate is in a position

vertical to the flow direction.

• The filter membrane module and the slide must be completely sub-

mersed (max. 40 cm).
Take changing water levels into account when mounting the filter
membrane module.

• The lower edge of the filter membrane module must be mounted

with least 10 cm distance to the bottom.

Installation

1

Mount the rail of the attachment for filtration in the basin. If nec-
essary, extend the rail with the
M-EXT 1.5 extension accessory.

Fig. 3-14: Sample filtration device (installed)

1 Chain (scope of delivery: Attachment for filtration M 1.5)
2 Guide rail (scope of delivery: Attachment for filtration M 1.5)
3 Height adjustable slide (scope of delivery: Suction line)
4 Suction line (scope of delivery: Suction line)
5 Sleeve tube (scope of delivery: Suction line)
6 Filter membrane module FM with membrane insert

2

Mount the filter membrane module on the slide of the attachment
for filtration.

3

Connect the suction line to the filter membrane module.

In special cases (e.g. in a channel) it is better to mount the
FM filter membrane module in a position horizontal to the
flow direction.

An adapter for horizontal mounting is available as an
accessory.

2

1

3

6

5

4

Advertising