4. Measuring / Operation

4.1 Measuring operation

4.1.1 Determination of measured values

The P 700 IQ determines the measured values with a chemical analyz-
ing procedure. Each measurement takes place in several steps.

1

Draw sample and rinse the measuring circle with sample

2

Measure the absorbance of the sample without reagent (blank
value)

3

Add the reagent

4

Leave to react for the reaction time

5

Measure the absorbance of the sample with reagent

The measured values are determined at intervals specified in the set-
tings (see section 4.1.4,  51).

If the intervals overlap with the specified start times and intervals of
other functions (Autom.cleaning or Autom.calibration), the functions
are carried out according to their priority (see section 4.1.4,  51).

4.1.2 Starting the measuring operation

Prior to starting the measuring operation, you always have to check the
readiness for operation of the individual components of the P 700 IQ.

The analyzer does not automatically start measuring when being con-
nected to the IQ S

ENSOR

N

ET

.

Start the measuring operation manually in the calibration and service
menu.

1

Using <>, select the P 700 IQ in the measured value display.

2

Using <C>, switch to the calibration and service menu.
The maintenance condition is activated.

3

Confirm the maintenance condition with <OK>.
The calibration and service menu is displayed.