4 calibration history – YSI P 700 IQ Analyzer User Manual

Page 61

Advertising
background image

ba76133e01

09/2013

61

P 700 IQ

20

Wait until a measured value is shown in the measured value dis-
play (approx. 5…7 minutes).

21

Switch off the maintenance condition.
See IQ S

ENSOR

N

ET

system operating manual.

Possible calibration

results

The calibration data are evaluated by the system. A calibration proce-
dure can have the following results:

4.2.4 Calibration history

Calibration history

Fig. 4-8: Calibration historyP 700 IQ

The calibration history provides the following information:

Date and time of the calibration

Type

M = manual calibration,
A = automatic calibration,
F = factory calibration

m (slope)

• * = 1-pt. cal (the slope was not determined)

no* = 2-pt. cal (the slope was determined)

Display after cali-
brating

Log book entries
(meaning/actions)

Measured value

Successful calibration

"----"

Sensor could not be calibrated.
Sensor blocked for measurement.
• Carry out maintenance activities imme-

diately
(see operating manual).

• View the calibration history.

• Check the calibration conditions and

calibration standard.

Currently active
calibration

Chronological list of
the last calibration
procedures