6. What to do if ...

Display “----”

Implausible measured

values

Cause

Remedy

No (valid) measured value avail-
able

Start measurement
and wait for the measurement to
be finished (5 .. 7 min)

Calibration error

Carry out calibration

Container for reagent empty

Check the liquids and connect a
new set of liquids if necessary

Liquid container not correctly
connected

Connect correctly

Sample intake does not work

Check the pump

Switch on the pump

Cleaning the overflow vessel

Power failure

Restart measurement in the cali-
bration and service menu.

Automatic start of measurement

Unknown

See log book

Cause

Remedy

Wait for another measurement

Calibration error

Carry out calibration

Measuring cell of the photometer
unit dirty

Carry out cleaning procedure

Container for reagent RE 2.5 is
empty

Exchange RE 2.5