Propaq MD Overview

9650-0807-01 Rev. A

Propaq MD Service Manual

7

Equipment You Need to Perform the Maintenance Tests

The equipment listed below is used in the maintenance procedures in this chapter. You can
substitute an equivalent device; however, not all simulators and analyzers will produce the
same results. Be sure to follow the manufacturer’s recommendations for conducting the
maintenance tests.

•

Fluke Impulse 7000DP Analyzer, with appropriate interface adapter (Fluke Biomedical

BP Pump/Dynatech Nevada Inc. part number 3010-0378)

•

ECG simulator

•

Fluke Index 2 SpO

2

simulator

•

IBP/Temp simulator

•

NIBP simulator

•

5% CO

2

calibration gas cylinder with CO

2

sample line (kit available from Scott Medical

Products, PN 4653ORF-BD)

•

CO

2

sampling line airway adapter

•

Stop watch

•

Battery Current Test Fixture 9100-3055-TF

•

Agilent E3633A Power Supply

•

Agilent 34401 Digital Multi-meter

Accessories You Need for the Propaq MD Options Maintenance Tests

•

3-lead, 5-lead and 12-lead ECG cables. (12-lead cable needed if 12-lead option is

installed.)

•

Dual lumen NIBP hose

•

IBP cable

•

SpO

2

cable

and sensor

•

EtCO

2

cable

and sensor

•

Paddles

•

Printer paper

•

Battery

•

Auxiliary power source

•

USB thumb drive