ZOLL Propaq MD Rev A User Manual

Page 59

Advertising
background image

Propaq MD Error Messages

9650-0807-01 Rev. A

Propaq MD Service Manual

55

NIBP TEMP OUT OF
RANGE

Module temperature is out of
its specified operating range.

Move the device to a cooler
or warmer environment.

NIBP WEAK PULSE

Too few pulses are detected
during the reading attempt.

Tighten the cuff and ensure
appropriate placement.

PACER CURRENT FAULT

Lack of pacer current.

Replace the defibrillator
board module.

PACER DEVICE FAILURE

Generation of pacer pulse
failed.

Replace the defibrillator
board module.

PRINTER OFFLINE

Printer module is off-line.

If it remains off line, try
checking the connectors or
power to the printer. If
problem persists, replace
the printer module.

PRINTER OUT OF PAPER

If there is paper in the
device, probably a paper
sensor problem.

Replace the printer module.

PRINTER PLATEN OPEN

If printer door is closed,
probably a problem with the
micro switch.

Replace the printer module.

PS DEFIB CHARGE
SHORT

One of the paddle set keys
are stuck.

Replace the paddle set or
MFC.

PS DEFIB ENERGY DOWN

Paddle set keys are stuck.

Replace the paddle set or
MFC.

PS DEFIB ENERGY DOWN
SHORT

Paddle set energy down
select button is stuck.

Replace the paddles.

PS DEFIB ENERGY UP

Paddle set keys are stuck.

Replace the paddle set or
MFC.

PS DEFIB FIRE SHORT

Paddle set keys are stuck.

Replace the paddle set or
MFC.

PS DEFIB RECORDER
SHORT

One of the paddle set keys
are stuck.

Replace the paddle set or
MFC.

QUICK ACCESS SW 1(to
SW 7) SHORT

One of the quick access
keys (key 1 to key 7) is
stuck.

Replace the front panel.

SCREEN CYCLE SHORT

Display/home key is stuck.

Replace the front panel.

Error Message

Explanation

Technical Action

Advertising