Ibp test – ZOLL Propaq MD Rev A User Manual
Page 39

Maintenance Tests
9650-0807-01 Rev. A
Propaq MD Service Manual
35
IBP Test
Tools Needed
IBP Simulator (with accuracy better than +/- 1 mmHg), or equivalent
Setup
Connect the test cables to IBP patient simulator.
Do This...
Observe This...
Pass/Fail
1.
Connect the IBP patient simulator to the
Propaq MD P1 channel.
2.
Set the IBP patient simulator to 0 mmHg.
3.
On the Propaq MD unit, press > IBP.
4.
On the Propaq MD unit, Press > P1->0<-. Display of 0/0(+/-2) in the P1 box
at the bottom of the display (P1 is
zeroed).
The waveform for channel 1 is
present.
o
o
o
o
5.
Exit the menu to see the waveform.
6.
Set the IBP patient simulator to 200
mmHg (or as close to 200 mmHg as your
simulator allows).
Display of 200/200(+/-5) in the P1
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P1
diastolic pressure is 200 (+/-5)).
o
o
7.
Repeat steps 1 through 5 of this procedure
for P2 and P3 channels.
8.
Disconnect IBP cable.