ZOLL Propaq MD Rev A User Manual

Page 58

Advertising
background image

C

HAPTER

2

T

ROUBLESHOOTING

54

9650-0807-01 Rev. A

NIBP KINKED HOSE

The cuff inflation was too
rapid. There may be kinked
tubing or another air
obstruction.

Check the external NIBP
hose for kinks. If problem
persists, check the routing
of the NIBP tubing inside
the device.

NIBP LOW VOLTAGE

Insufficient voltage is
supplied to the NIBP
module.

Check the module voltage
in the NIBP Test screen.
Inspect the unit or water
ingress. If evidence of
water ingress is found,
allow the water to
evaporate while periodically
checking the module
voltage. If problem persists,
replace the NIBP module.

NIBP MAX PRESSURE
EXCEEDED

Pressure limit exceeded.

If problem persists, do the
following:

Check the calibration via
the “NIBP Calibration
Check” on page 36.

Check the routing of the
NIBP tubing inside the
device for kinks or
obstructions.

If the NIBP module is out of
calibration, replace the
NIBP module.

NIBP MONITORING
FAILURE

One of the following issues
occurred:

A failure was reported by
the NIBP module.

The two transducers are
not aligned. They are
either not calibrated,
there is a transducer
failure, or there is a
kinked tube.

A calibration problem was
reported by the NIBP
module.

If problem persists, replace
the NIBP module.

NIBP NEONATAL
DETECTED

Device thinks a neonate cuff
is connected. If no neonate
cuff is connected, tubing
may be damaged or kinked.

Verify that neonate cuff is
not connected. If problem
persists, check the routing
of the NIBP tubing inside
the device.

Error Message

Explanation

Technical Action

Advertising