ZOLL Propaq MD Rev A User Manual

Page 82

Advertising
background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

78

9650-0807-01 Rev. A

Removing the Defibrillator Boards and Capacitor Bank
Modules

Tools Required

#1 Phillips screwdriver with torque capability

#2 Phillips screwdriver with torque capability

26 mm open end wrench

Large diagonal cutters

Setup

This procedure should be conducted on a static mat.

Wear a wrist strap to ground during the procedure.

Complete these procedures in the following order before performing this procedure:

1.

“Removing the Handle” on page 61

2.

“Removing the Printer Module” on page 62

3.

“Disassembling the Main Unit” on page 63

To reinstall the defibrillator boards and/or capacitor bank modules, reverse
the steps.

Defibrillator Boards and Capacitor Bank Disassembly

1

Lay back end of separated unit down with
opened side facing up.

2

Pull the black snap rivet straight out.

3

Disconnect the defib-system flex cable
from the processor board.

4

Pull out small interconnect board.

Advertising